Drilling in Two Steps, contd.

Click on image for enlarged view
Remove the insert drill sleeve 3.   Remove the insert drill sleeve. Using the Depth Gauge for locking bolts, measure for the bolt. Select a locking bolt 2 mm longer than the measured length to compensate for the tapered trocar tip and ensure full engagement of the far cortex.


Next Page

UNS Index