Distal Locking with the Distal Aiming Device, contd.
Click on image for enlarged view
9. Make a stab incision over the distal hole. Insert the 11.0 mm/8.0 mm
Protection Sleeve and 8.0 mm Trocar through the Direction Finder
and stab incision onto the bone. Replace the trocar with the 8.0 mm/4.5 mm Drill Sleeve. Drill through
both cortices using the 4.0 mm/4.5 mm Drill Bit.
Remove the drill sleeve. Measure for the proper length locking bolt.
Add 2 mm to the measurement to ensure full engagement of the far
cortex. Insert the locking bolt through the 11.0 mm/8.0 mm Protection Sleeve.