Distal Locking with the Distal Aiming Device, contd.

image intensification required image intensification required for this step image intensification required

Click on image for enlarged view
Shift the aiming device over the bone
4.   Push the Distal Aiming Device firmly against the bone surface. Remove the Aiming Trocar and replace it with the 8.0 mm/4.5 mm Drill Sleeve. Confirm the positioning with the Direction Finder. If the aiming device is correctly aligned, the locking hole in the nail appears round through the drill sleeve, and the dot is in the center of the Direction Finder.