Proximal Schanz Screw, contd.

Make a 1 cm-2 cm skin incision, and bluntly dissect the soft tissue down to the bone. Insert the triple trocar assembly through the Curved Aiming Attachment and onto the bone. This should be at the level of the lesser trochanter. Remove the trocar. Using the Oscillating Attachment and a 3.5 mm Three-Fluted Drill Bit, drill through both cortices. Remove the 5.0 mm/3.5 mm Drill Sleeve.

Using the Universal Chuck with T-Handle, insert a 5.0 mm Schanz Screw into the calcar. Remove the 6.0 mm/5.0 mm Drill Sleeve and Curved Aiming Attachment.



Next Page