Distal Locking with the Radiolucent Drive, contd.

   Click on image for enlarged view
Distal Locking with the Radiolucent Drive
 - center drill bit in locking hole
image intensification required image intensification required for this step image intensification required

3.   Insert the special 4.0 mm Three-Fluted Drill Bit [511.46] into the Radiolucent Drive. Under image intensification, place the tip of the drill bit oblique to the X-ray beam, into the stab incision and onto the femur, until the tip of the drill bit is centered in the locking hole (Fig. C).