Distal Locking with the Radiolucent Drive, contd.
image intensification required for this step
Click on image for enlarged view
3. Insert the special 4.0 mm Three-Fluted Drill Bit [511.46] into the Radiolucent Drive. Under image
intensification, place the tip of the drill bit oblique to the X-ray beam, into the stab incision and onto the
tibia, until the tip of the drill bit is centered in the locking hole image
(Fig. C).