Locking Proximally, contd
Attach the Aiming Arm
Click on images for enlarged views
For transverse locking, orient the
aiming arm for medial to lateral bolt
insertion. For interlocking through
the oblique locking hole, attach the
aiming arm to the insertion handle
for an anterolateral or anteromedial
approach. Tighten the spring-loaded
connecting knob to secure the aiming
arm to the insertion handle. Confirm
that the insertion handle is securely
fastened to the nail. Tighten the
insertion bolt if necessary.
Next Page