Inserting the Nail

Assemble the Insertion Instruments

Assemble the Insertion Instruments
Click on image for enlarged view

Orient the insertion handle anteriorly, and match the flats on the handle and nail.

For solid nails, select the Insertion Bolt for Titanium Solid Tibial Nails [356.542]. For cannulated nails, select the Insertion Bolt for Titanium Cannulated Tibial Nails [356.544].

Place the insertion bolt into the insertion handle and thread it into the proximal nail end. Tighten the insertion bolt with the Cannulated Socket Wrench [355.14], but do not over tighten.

Thread the inserter-extractor onto the threaded post of the insertion handle.

Note:
Do not attach the aiming arm for nail insertion.